SP844
Physiological Pressure Transducer Specifications
|
| Pressure
Range:
|
-20 to 300 mmHg
|
| Max.
Overpressure:
|
10000 mmHg
|
| Sensitivity:
|
50 µV/V/cm Hg
|
| Resonance
Frequency *:
|
300 Hz typical (transducer and dome)
|
| Max.
Electrical Excitation:
|
15 V DC or AC
|
| Input
Resistance: |
700 W (or
equipment specified)
|
| Output
Resistance: |
1000 W (or
equipment specified)
|
| Non-Linearity
and Hysteresis:
|
max. ±0.5 % of full scale
|
| Zero
Balance:
|
max. + 30 mmHg
|
| Thermal
Sensitivity Shift: |
0.15 %/oC
|
| Thermal
Zero Shift: |
max.: 0.25 mmHg/oC
|
| Operating
Temperature Range: |
+ 10 to + 50oC
|
| Storage
Temperature Range: |
- 20oCto
+ 70oC
|
| Insulation
Resistance: |
minimum 10MW
|
| Leakage
Current:
|
max. 1.5 µA at 250 V, 50Hz
|
| High
Voltage Resistance:
|
10000 V
between saline in dome
and transducer leads
|
| Weight:
|
24 grams (without cable)
|
| Length
of Cable:
|
0.3 m or customer specified
|
| Adapter
Cable:
|
2.5 m with equipment specified
connector or open-ended
|
*
The frequency response may be limited by pressure tubing.
Specifications may be subject to change without notice. |